chemical
pH & Alkalinity
Purpose
Measure the hydrogen-ion concentration (pH) and the alkalinity of the mud and filtrate to assess lime content, bicarbonate/carbonate balance, and overall system buffering capacity. These values guide treatment decisions for water-based fluids.
Equipment
- pH meter with temperature compensation, or pH paper (wide range, then narrow range)
- Buffer solutions for pH calibration (pH 7 and pH 10 typical)
- Phenolphthalein indicator solution
- Methyl-orange indicator solution (for Mf)
- Titration burette or syringe (0.01 N H₂SO₄)
- Filtrate sample (from API filtration test)
- Mud sample
- White porcelain spot plate
Procedure
pH (meter method):
- Calibrate the pH meter with buffer solutions according to the manufacturer’s instructions.
- Rinse the electrode with distilled water and blot dry.
- Immerse the electrode in the mud sample (or filtrate) and allow the reading to stabilise.
- Record the pH value and sample temperature.
Pm — Mud Alkalinity (phenolphthalein):
- Place 1 mL of mud on the spot plate or in a small beaker.
- Add 2–3 drops of phenolphthalein indicator.
- Titrate with 0.01 N H₂SO₄ until the pink colour disappears.
- Record the volume of acid used (mL). Pm = mL of acid used.
Pf — Filtrate Phenolphthalein Alkalinity:
- Place 1 mL of filtrate in a small beaker; add 2–3 drops of phenolphthalein.
- Titrate with 0.01 N H₂SO₄ to colour change. Record Pf = mL acid.
Mf — Filtrate Methyl-Orange Alkalinity:
- Continue titrating the Pf sample, or use a fresh 1 mL filtrate aliquot.
- Add 2–3 drops of methyl-orange indicator.
- Titrate with 0.01 N H₂SO₄ from blue/green to orange endpoint. Record Mf = total mL acid.
Notes
- Report pH, Pm, Pf, and Mf on the daily mud report.
- pH paper may be used for quick field checks but the meter method is preferred for accuracy.
- Calibrate the pH meter with fresh buffer solutions at the start of each shift.
- High Pm relative to Pf indicates significant excess lime in the mud phase.